Poster | 247th ECS Meeting | Montreal, QC
Daniela Vieira, Emad Shahnam, Jared Lapkovsky
Delong America, Montreal, QC, Canada
At the 247th ECS Meeting in Montreal, we showed how low voltage electron microscopy (LVEM) at 5 and 25 kV images carbon nanotubes, carbon dots and graphene sheets with no observed beam damage and a broader contrast range than a 200 kV TEM.
Carbon-based materials play an important role in electrochemical systems. Their large surface area, many active sites and wide potential window make them relevant to energy storage and sensing. Porous carbon, carbon dots, graphene and carbon nanotubes are all widely used, either pristine or functionalized with other materials through covalent bonding.
Transmission electron microscopy (TEM) has been a key technique for understanding carbon morphology, structural changes and integrity since the 1940s. Even though modern TEMs reach resolutions down to 0.05 nm, beam damage at high accelerating voltages (80 kV and above) remains a challenge because the bonds between light elements such as B, C, N and O are fragile. A common recommendation is to expose analysis areas for very short periods (under 15 s) at beam energies limited to 100 kV.
Lowering the accelerating voltage reduces both knock-on damage and radiolysis of carbon-based materials. LVEM also offers higher contrast than high voltage TEM (HV-TEM) for light elements, and is more compact and cost-effective than a traditional HV-TEM.
Key findings
Solutions of dispersed carbon nanotubes (CNTs), graphene and carbon dots (CDs) were ultrasonicated and drop cast onto carbon-coated copper grids. Images were acquired with the LVEM 5 at 5 kV and the LVEM 25E at 25 kV (Delong Instruments). The graphene sample was also imaged with an HV-TEM at 200 kV to compare image contrast, and grayscale profiles were measured in ImageJ. Electron diffraction (ED) was recorded on the graphene sheets.

Contrast analysis of graphene TEM images at (A) 25 kV on the LVEM 25E and (B) 200 kV on an HV-TEM, with (C) the grayscale profiles of both images.

Carbon nanotubes, LVEM 5 at 5 kV.

Carbon nanotubes, LVEM 25E at 25 kV.

Carbon dots, 8 to 10 nm in diameter, LVEM 25E at 25 kV.

Graphene sheets, LVEM 5 at 5 kV.
Scale bars are shown on each image.
The typical structure of the carbon nanotubes was clearly observed, with a diameter of approximately 10 nm. Carbon dots measured 8 to 10 nm in diameter. The morphology and size of all the carbon nanomaterials were clearly resolved and were consistent with HV-TEM data.
Because of the low accelerating voltage, no beam damage was observed, which allowed prolonged exposure to improve data acquisition. Images at 25 kV were sharper than those at 5 kV, yet both voltages gave sufficient resolution to characterize the carbon nanoparticles. The electron diffraction pattern of the graphene sheets showed the expected set of hexagonal diffraction spots, consistent with single crystal graphene.
For contrast, the grayscale range of the graphene images was compared. The 25 kV LVEM image spanned gray values from 60 to 250, while the 200 kV HV-TEM image spanned only 50 to 125. As expected, the lower accelerating voltage gives better contrast for low atomic number elements because they scatter electrons more strongly. This is a significant advantage when imaging carbon nanotubes, carbon dots and graphene.
Beyond contrast and reduced beam damage, LVEM brings a lower initial cost, lower operating cost, easier operation and maintenance, a smaller laboratory footprint and no need for specialized site preparation.
Low voltage TEM is a powerful tool for characterizing the morphology, size and structure of carbon-based materials, offering more contrast than traditional HV-TEM. It is also more affordable, easier to use, and needs less specialized infrastructure and training.