Poster | Microscopy & Microanalysis 2025 | Salt Lake City, UT

Hydrophilization and Activation of Carbon-Coated TEM Grids Using a Light Spectrum Device

Daniela Vieira, M. Safari, H. Hosseinkhannazer, Emad Shahnam, E. West, Jared Lapkovsky

Delong America, Montreal, QC, Canada | Norcada, AB, Canada

In a joint study with Norcada presented at Microscopy & Microanalysis 2025, the LVEM 25E was used to evaluate UltiFlow, a compact light-based grid treatment, as a vacuum-free alternative to glow discharge for preparing hydrophilic carbon-coated TEM grids.

The quality of any TEM analysis depends on how well the specimen is deposited on the grid. Nanoparticles are usually prepared by drop casting: the particles are dispersed by ultrasonication, a droplet is placed on the grid, the excess is blotted away and the grid is left to dry. As the solvent evaporates, surface tension pulls particles together, and the resulting agglomerates make it hard to resolve and measure individual particles.

Glow discharge is the standard fix. Plasma generated under vacuum reacts with the carbon film and renders it hydrophilic. However, commercial glow discharge systems need a vacuum pump and regular maintenance, typically weigh over 20 kg, and operate in a narrow process window where a few extra seconds can damage the grid.

Norcada's UltiFlow takes a different route. It uses a broad-spectrum light source to activate and hydrophilize the carbon surface within minutes, with the effect lasting up to one hour after treatment. It needs no vacuum, weighs under 2 kg and is maintenance-free. This work optimized the UltiFlow treatment and compared it with glow discharge and untreated grids.

Key findings

  • A 5 minute treatment at 50% power was identified as the optimal UltiFlow setting, balancing hydrophilization with carbon film integrity.
  • The strongest hydrophilic effect (contact angle 72.8° ± 4.2) was reached at 10 minutes and 30% power, versus 103.0° ± 1.4 on untreated grids.
  • UltiFlow-treated grids gave well-distributed gold nanoparticles comparable to glow discharge, and a more homogeneous distribution of polymer particles.
  • LVEM 25E imaging at 25 kV was used to verify grid integrity and particle dispersion from predefined, fixed grid positions.

Methodology

Low-voltage electron microscopy (LVEM 25E, Delong Instruments) was used to characterize the carbon film after light treatment and to quantify particle distribution and homogeneity. Images were acquired at 25 kV from predefined, fixed areas of each grid using the microscope's software-controlled movement panel, so that treatments could be compared on equivalent fields.

Hydrophilicity was measured by contact angle: a 10 µL water droplet was placed on untreated and treated grids and the angle was determined in ImageJ, with values below 90° indicating a hydrophilic surface. Gold nanoparticles (AuNPs), polymer particles and carbon nanotubes (CNTs) were dispersed in water, ultrasonicated for 15 minutes and drop cast onto untreated, UltiFlow-treated or glow discharge-treated grids within 30 minutes of pre-treatment.

Results Imaged on the LVEM 25E

Contact angle droplets and LVEM images of TEM grid integrity after UltiFlow light treatment at different times and power levels

Fig A. Contact angle and grid integrity after light treatment at varying exposure times and power intensities. Carbon film damage appears at 100% power and longer exposures.

LVEM 25E TEM images of gold nanoparticles, polymer particles and carbon nanotubes on untreated, UltiFlow and glow discharge treated grids

Fig B. Gold nanoparticles, polymer particles and carbon nanotubes deposited on untreated, UltiFlow-treated and glow discharge-treated grids.

TEM images acquired with the LVEM 25E at 25 kV.

Results and Discussion

Exposure time and power were varied (1, 5 and 10 minutes; 30%, 50% and 100%). Short or low-power treatments left the contact angle essentially unchanged from untreated grids (103.0° ± 1.4 untreated, 103.8° ± 1.4 at 1 minute and 100%, 106.0° ± 2.2 at 5 minutes and 50%). A modest improvement appeared at 5 minutes and 100% (95.0° ± 2.8), and the most pronounced effect at 10 minutes and 30% (72.8° ± 4.2).

LVEM imaging showed carbon film damage at the highest power and longest exposures, with 10 minutes at 100% visibly damaging the film. Weighing hydrophilization against integrity, 5 minutes at 50% was selected as the optimal UltiFlow setting for most applications.

On untreated grids, gold nanoparticles formed large agglomerates. UltiFlow-treated grids gave well-dispersed, minimally agglomerated AuNPs, comparable to glow discharge, and achieved an even more homogeneous distribution of polymer particles than glow discharge. For CNTs, no significant difference was seen between treatments, likely because of the inherently high surface energy and surface area of nanotubes.

UltiFlow Compared With Conventional Glow Discharge

Both approaches render carbon-coated grids hydrophilic in about five minutes. The practical differences lie in the hardware a lab needs to run them.

Conventional glow discharge

  1. Vacuum pump required (oil or oil-free)
  2. About 20 kg device plus pump, not portable
  3. 25 to 81 grids per treatment
  4. Regular oil changes and maintenance

Higher cost, narrow process window

Norcada UltiFlow

  1. No vacuum pump needed
  2. Under 2 kg and portable
  3. Up to 12 grids per treatment
  4. Maintenance-free

Lower cost, compact and simple

Norcada UltiFlow light-based TEM grid treatment device
Delong Instruments LVEM 25E low-voltage electron microscope

The Norcada UltiFlow (left) and the LVEM 25E (right): a space-saving pairing for routine nanoparticle sample preparation and imaging.

Conclusion

The light-based surface treatment provided by Norcada UltiFlow is a suitable and effective alternative to glow discharge for converting carbon-coated TEM grids from hydrophobic to hydrophilic. Combining a broad-spectrum light source with controlled heating around the grids, it minimized particle agglomeration and enabled reliable nanoparticle characterization.

LVEM provided rapid, high-contrast imaging across a wide range of nanoparticle types. Together, the compact UltiFlow and the LVEM 25E offer a space-saving, cost-effective and efficient workflow for routine nanoparticle analysis.

Presented at

Microscopy & Microanalysis 2025

Joint poster with Norcada | July 27 to 31, 2025 | Salt Lake City, UT

Book a Demo

References

  1. Mulligan SK, et al. Multiplexed TEM specimen preparation and analysis of plasmonic nanoparticles. Microscopy and Microanalysis. 2015;21(4):1017-1025. doi:10.1017/S1431927615014324
  2. Michen B, et al. Avoiding drying-artifacts in transmission electron microscopy: characterizing the size and colloidal state of nanoparticles. Scientific Reports. 2015;5:9793. doi:10.1038/srep09793
  3. Zheng J. Improving sample preparation methods to assess nanoparticle agglomeration using TEM. Microscopy and Microanalysis. 2014;20(S3):1236-1237. doi:10.1017/S1431927614007910
  4. Kaliyaraj Selva Kumar A, Zhang Y, Li D, Compton RG. A mini-review: how reliable is the drop casting technique? Electrochemistry Communications. 2020;121:106867. doi:10.1016/j.elecom.2020.106867
  5. Kazan R, et al. Grid batch-dependent tuning of glow discharge parameters. Frontiers in Molecular Biosciences. 2022;9. doi:10.3389/fmolb.2022.910218
  6. Aebi U, Pollard TD. A glow discharge unit to render electron microscope grids and other surfaces hydrophilic. Journal of Electron Microscopy Technique. 1987;7(1):29-33. doi:10.1002/jemt.1060070104
  7. Norcada UltiFlow. norcada.com/products/norcada-ultiflow
  8. Kumar S. Investigating effect of CNT agglomeration in CNT/polymer nanocomposites using multiscale finite element method. Mechanics of Materials. 2023;183:104706. doi:10.1016/j.mechmat.2023.104706

Get the full poster

Free PDF, sent straight to your inbox. Four quick details and it is on its way.

M&M 2025 poster: Hydrophilization and activation of carbon coated TEM grids using a light spectrum device
Please enter your first name

Please enter your last name

Please enter a valid phone number

Please enter a valid email address

Please enter an institution name.